.. index:: pair: Gamma; Settings .. _gammasettings: Gamma ===== These contain customisable options for the :ref:`Gamma`. Available settings are: * **Distance to agreement**: Distance-to-agreement in mm; e.g. 2 mm. * **Dose threshold**: The dose threshold percentage of the maximum dose, below which is not analyzed. Must be between 0 and 1. * **Dose to agreement**: Dose-to-agreement in percent; e.g. 2 is 2%. * **Gamma cap**: Limits the range of displayed Gamma values. * **Global dose**: True if the global dose maximum is to be used, otherwise the local dose value is used. |Note| Not all settings have been implemented. .. |Note| image:: _static/Note.png