.. index:: pair: Starshot; Settings .. _starshotsettings: Starshot ======== These contain customisable options for the :ref:`starshot`. Available settings are: * **DPI**: Image resolution in Dots Per Inch (DPI). Usually needed for scanned film. * **Normalised analysis radius**: Distance in % between starting point and closest image edge; used to build the circular profile which finds the radiation lines. Must be between 0.05 and 0.95. * **Recursive analysis**: If True will recursively search for a “reasonable” wobble. * **SID**: Source Image Distance (SID) in mm. Usually needed for scanned film. * **Tolerance**: The tolerance in mm to test against for a pass/fail result. |Note| Not all settings have been implemented. |Hint| :ref:`Inverting` the image sets the invert flag. Invert the image if the automatically-determined pylinac inversion is incorrect. .. |Note| image:: _static/Note.png .. |Hint| image:: _static/Hint.png