Picket FenceΒΆ

These contain customisable options for the Picket Fence Analysis. Available settings are:

  • Apply median filter: If true applies a median filter of size 3.

  • Leaf Action: Defines an action level in mm for leaf pair deviation. Must be less than Leaf Tolerance.

  • Leaf Tolerance: The tolerance of difference in mm between an MLC pair position and the picket fit line.

  • MLC Type: Sets the default Multi-Leaf Collimator (MLC) the image was acquired from. Select from the list provided.

  • Number of pickets: The number of pickets in the image.

Note Not all settings have been implemented.

Hint Inverting the image sets the invert flag.