Tomographic ContrastΒΆ

These contain customisable options for the Tomographic Contrast Analysis. Available settings are:

  • Search Slices: Number of image slices to search from sphere starting points

  • Search window px: Number of pixels to search from sphere starting points

  • Sphere angles: Comma delimited list of sphere angles starting at horizontal right and going counterclockwise.

  • Sphere diameters mm: Comma delimited list of sphere diameters in mm corresponding to the sphere angles

  • UFOV ratio: Defines the Useful Field Of View (UFOV) over the phantom as a ratio of the phantom outline.